Dental Burs and Endodontic Files: Are Routine Sterilization Procedures Effective?FULL TEXT
• Archie Morrison, DDS, MS, FRCD(C) • A b s t r a c tMaterials and Methods: The sterility of new (unused) and used dental burs and endodontic files before and after various sterilization procedures was analyzed. New burs and files were tested immediately after removal from manufacturers' packaging, with or without prior sterilization. Burs and files that had been used in various dental offices were precleaned, packaged, resterilized and then tested for various pathogens. Each item was individually removed from the sterilization packaging, transferred by sterile technique into Todd-Hewitt broth, incubated at 37°C for 72 hours and observed for bacterial growth. Results: Sterilization procedures were 100% effective for unused burs and unused files but were less than 100% effective for all other test groups. Contamination rates following sterilization ranged from 15% for one group of used burs (p = 0.01) to 58% for one group of used files (p < 0.001). Conclusions: Dental burs and endodontic files, as packaged by the manufacturer, are not sterile and should therefore be sterilized before first use. The resterilization procedures tested here were not adequate, and more rigorous sterilization procedures are needed. If such procedures cannot be devised, these instruments should perhaps be considered single-use devices.
|
|
Full text provided in PDF format |
|